GPR-1100
Portable ppm Oxygen Analyzer
Owner’s Manual
2855 Metropolitan Place, Pomona, CA 91767 USA Tel: 909-392-6900, Fax: 909-392-3665, e-mail: [email protected], www.aii1.com Rev 7/06
Przeglądanie stron 0
1 2 3 4 5 6 ... 34 35

Podsumowanie treści

Strona 1

GPR-1100 Portable ppm Oxygen Analyzer Owner’s Manual 2855 Metropolitan Place, Pomona, CA 91767 USA ♦ Tel: 909-392-6900, Fax

Strona 2 - Table of Contents

10 Calibration & Accuracy Single Point Calibration: As previously described the galvanic oxygen sensor generates an electrical current sen

Strona 3 - 1 Introduction

11 Example: As illustrated by Graph A any error, play in the multi-turn span pot or the temperature compensation circuit, during a span a

Strona 4

12 Start-up The GPR-1100 Portable ppm Oxygen Analyzer is fully operational from the shipping container with the oxygen sensor installed and calib

Strona 5 - 3 Safety

13 Gas Connections: The GPR-1100 flow through configuration is designed for positive pressure samples and requires connections to incoming sample

Strona 6 - Maintenance

14 Power connection: Locate a source of AC power to meet the area classification, plug in the appropriate charging adapter to the outlet and conn

Strona 7 - 5 Operation

15 Establishing Power to the Electronics: Establish power to the analyzer electronics by pushing the red ON/OFF key. The digital display responds

Strona 8 - Pressure & Flow

16 Main Menu: Access the MAIN MENU by pressing the MENU key: MAIN MENU AUTO SAMPLE MANUAL SAMPLE SPAN CALIBRATE ZERO CALIBRATE DEFAULT SPAN DE

Strona 9

17 For example, if the analyzer is reading 1% on the 0-10% range and an upset occurs, the display will shift to the 0-25% range when the oxygen re

Strona 10

18 0.000 PPM MANUAL SAMPLING 1000 PPM RANGE 24.5 C OR 0.000 PPM OVERRANGE AUTO SAMPLING 1000 PPM RANGE

Strona 11

19 Zero Calibration In theory, the oxygen sensor produces no signal output when exposed to an oxygen free sample gas. However, the analyzer will g

Strona 12 - Start-up

2 Table of Contents Introduction 1 Quality Control Certification 2 Safety 3 Features & Specifications 4 Operation 5 Maintenance 6 Spar

Strona 13

20 Both the Zero Calibrate and Span Calibrate functions result in the following displays: PASSED CALIBRATION OR FAILED CALIBRATION

Strona 14

21 Manual Span The user must ascertain that the oxygen reading (actually the sensor’s signal output) has reached a stable value within the limits

Strona 15

22 5. Assure there are no restrictions in the span gas line. 6. Assure the male fitting designated for the vent is installed and the vent is ope

Strona 16

23 21. Both the Zero Calibrate and Span Calibrate functions result in the following displays: PASSED CALIBRATION OR FAILED CALIBRA

Strona 17

24 8. Caution: If the analyzer is equipped with an optional sampling pump and is intended for use in both positive and atmospheric/slightly negat

Strona 18

25 Standby ¾ The analyzer has no special storage requirements. ¾ The sensor should remain connected during storage periods. ¾ Store the analy

Strona 19 - Zero Calibration

26 6 Maintenance Generally, cleaning the electrical contacts or replacing filter elements is the extent of the maintenance requirements of this a

Strona 20 - Span Calibration

27 7 Spare Parts Recommended spare parts for the GPR-1100 Series Portable Oxygen Analyzer: Item No. Description GPR-12-333 ppm Oxygen Sen

Strona 21

28 8 Troubleshooting Symptom Possible Cause Recommended Action Slow recovery At installation, defective sensor Air leak in samp

Strona 22

29 Symptom Possible Cause Recommended Action O2 reading doesn’t agree to expected O2 values Pressure and temperature of the sample is differen

Strona 23 - Sampling

3 1 Introduction Your new portable oxygen analyzer incorporates an advanced electrochemical sensor specific to oxygen along with state-of-the-art

Strona 24 - Battery Considerations

30 Symptom Possible Cause Recommended Action Erratic O2 reading or Negative O2 reading or No O2 reading p

Strona 25 - Standby

31 9 Warranty Policy The design and manufacture of Advanced Instruments oxygen analyzers and oxygen sensors are performed under a certified Quali

Strona 26 - 6 Maintenance

32 10 Material Safety Data Sheet ( MSDS ) Product Identification Product Name Oxygen Sensor Models GPR, PSR Series Synonyms Galvanic Fuel Cell

Strona 27 - 7 Spare Parts

33 Reactivity Data Stability Stable Conditions Contributing to Instability None Incompatibility Avoid contact with strong acids Hazardous Decomp

Strona 28 - 8 Troubleshooting

34 Product Identification Product Name Oxygen Sensor Models XLT Synonyms Galvanic Fuel Cell, Electrochemical Transducer Manufacturer Analytical

Strona 29

35 Incompatibility Avoid contact with strong bases Hazardous Decomposition Products Emits toxic fumes when heated Conditions to Avoid Heat Spil

Strona 30

4 2 Quality Control Certification Date: Customer: Order No.: Pass Model GPR-110

Strona 31 - 9 Warranty Policy

5 3 Safety General This section summarizes the essential precautions applicable to the GPR-1100 Series Portable ppm Oxygen Analyzer. Additional p

Strona 32

6 Maintenance Serviceability: Except for replacing the oxygen sensor, there are no parts inside the analyzer for the operator to service. Only tra

Strona 33

7 5 Operation Principle of Operation The GPR-1100 portable oxygen analyzer incorporates a variety of ppm range advanced galvanic fuel cell type

Strona 34

8 Pressure & Flow All electrochemical oxygen sensors respond to partial pressure changes in oxygen. The inlet pressure must always be higher

Strona 35

9 Application Pressure - Atmospheric or Slightly Negative: For accurate ppm range oxygen measurements, an optional external sampling pump should b