Owner’s Manual
GPR-3100
Oxygen Purity Analyzer
Advanced Sensor Technology
24 Month Operating Life in 100% Oxygen
Inherent Accuracy 0.1%, Resolution 00.01%
Auto Ranging & Suppressed Range Capability
Temperature Controlled & Barometric Compensationsy
Przeglądanie stron 0
1 2 3 4 5 6 ... 44 45

Podsumowanie treści

Strona 1 - Owner’s Manual

Advanced Instruments Inc. 2855 Metropolitan Place, Pomona, CA 91767 USA ♦ Tel: 909-392-6900, Fax: 909-392-3665, e-mail: info@aii2.

Strona 2

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 3 Sample System: The GPR-3100 is supplied with sta

Strona 3

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 4 throttle valve on the pump’s inlet side may be n

Strona 4

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 5 Calibration and Accuracy Single Point Calibratio

Strona 5

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 6 Graph C illustrates these 'worse case'

Strona 6

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 7 Installation The GPR-3100 Oxygen Analyzer consis

Strona 7

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 8 Mounting the Analyzer: The GPR-3100’s 10.8”W x 7

Strona 8

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 9 Gas Connections: The GPR-3100 is designed for po

Strona 9

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 10 Procedure: Caution: Do not change the factory s

Strona 10 - 5 Operation

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 11 Output Connections As illustrated above the sen

Strona 11

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 12 Power Failure Alarm A dry contact rated at 30VD

Strona 12

Advanced Instruments Inc. Table of Contents Introduction 1 Quality Control Certification 2 Safety 3 Features & Specifications 4 Operation

Strona 13

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 13 Installing the Oxygen Sensor The GPR-3100 is eq

Strona 14

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 14 Installing Span Gas Caution: Do not contaminat

Strona 15

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 15 Establishing Power to the Electronics: Once the

Strona 16

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 16 Operating the Menu Driven Controls – The appen

Strona 17

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 17 Range Selection: The analyzer display defaults

Strona 18

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 18 4. Press the ENTER key when F-C starts flashin

Strona 19

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 19 Calibration - General Guideline Zeroing the ana

Strona 20

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 20 consume the oxygen that has dissolved into the

Strona 21

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 21 span calibration again for up to 3 months under

Strona 22

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 22 16. The analyzer returns to the SAMPLE mode aft

Strona 23

Advanced Instruments Inc. 1 Introduction Thank You For Your Business Your new oxygen analyzer is a precision piece of equipment designed to g

Strona 24

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 23 To avoid damaging the sensor or sample system o

Strona 25

Advanced Instruments Inc. 6 Maintenance GPR-3100 Rev 7/2004 6

Strona 26

Advanced Instruments Inc. 7 Spare Parts Recommended spare parts for the GPR-3100 % High Purity O2 Analyzer include: Item No. DescriptionGPR-11

Strona 27

Advanced Instruments Inc. 8 Troubleshooting Symptom Possible Cause Recommended Action Reading does not reflect expected values Sensor was not c

Strona 28

Advanced Instruments Inc. 9 Warranty What is covered: Any defect in material and workmanship from normal use in accordance with the Owner’s Manu

Strona 29

Advanced Instruments Inc. 9 Warranty What is not covered: This warranty does not cover installation; defects resulting from accidents; damage wh

Strona 30

Advanced Instruments Inc. 10 Material Safety Data Sheet (MSDS) Product Identification Product Name Oxygen Sensor Models CAD, GPR, PSR, SAF, 670

Strona 31

Advanced Instruments Inc. 10 Material Safety Data Sheet (MSDS) Fire and Explosion Data Flash and Fire Points Not applicable Flammable Limits N

Strona 32

Advanced Instruments Inc. 10 Material Safety Data Sheet (MSDS) Symptoms Eye contact - burning sensation. Skin contact - soapy slick feeling. Me

Strona 33

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Strona 34

Advanced Instruments Inc. 2 Quality Control Certification Date: Customer: Order No.: Pass Mo

Strona 35

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Strona 36

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Strona 37

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Strona 38

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Strona 39 - 99.55 %

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Strona 40 - 99.55 % 99.55 %

Advanced Instrtuments Inc.5 Operation - Appendix A* MAIN MENU Sample Sample Span Zero Alarm SystemAuto Range 90 t

Strona 41 - 00.05 %

Advanced Instruments Inc. 3 Safety This section summarizes the essential generic precautions applicable to all analyzers. Additional precautions

Strona 42 - 5 Operation - Appendix A

Advanced Instruments Inc. 3 Safety Serviceability: Except for replacing the oxygen sensor, there are no parts inside the analyzer for the operato

Strona 43

Technical Specifications Accuracy: < 1% of FS range at constant conditions; inherent accu-racy of 0.1% following calibration with 95-100% oxygen

Strona 44

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 1 Principle of Operation The GPR-3100 oxygen analy

Strona 45

Advanced Instruments Inc. 5 Operation GPR-3100 Rev 2/05 2 With the sensor and sample gas lines temperature